FALCON HANDPIECE CLEANING, MAINTENANCE; OVERHAUL
EN
INTELECT
®
RPW 2 USER MANUAL
FALCON
®
HANDPIECE CLEANING, MAINTENANCE; OVERHAUL
CLEANING
Regular cleaning ensures perfect hygiene and operation of the Falcon
®
handpiece. The handpiece, in particular the shock
transmitter, must be thoroughly cleaned and disinfected after each therapy session.
Clean the coupling gel o the handpiece using a cleaning agent that is suitable for surface cleaning.
Disinfect the handpiece with an alcohol-based disinfectant that is suitable for surface cleaning.
Component Procedure Interval
Handpiece shaft and cushion clean and disinfect
daily
or after , shocks (whichever comes first)
Guide tube clean from inside with brush every day
Shock transmitters and O-rings clean in ultrasonic bath and disinfect after each treatment or contact with a patient
Guide tube, projectile and O-rings replace after ,, shocks (overhaul the handpiece)
Caution!
Flammable and volatile cleaning and disinfecting agents may generate an explosive atmosphere.
Disconnect the handpiece from the control unit before starting any cleaning or maintenance work.
Attention!
It is essential that no fluid be permitted to penetrate either the device or its tubing.
*Falcon is a registered trade mark of Storz medical AG