85
CLEANING, MAINTENANCE, OVERHAUL
EN
INTELECT
®
RPW 2 USER MANUAL
. Dispose of the used guide tube and the used projectile
. Dispose of the detachable sealing rings of the C, DI,
F, B, T shock transmitters and the sealing ring on the
shaft.
. Clean the shaft, the shock transmitter (including firmly
seated sealing rings) and the shock transmitter screw cap
using a disinfectant containing alcohol. These are reused
after cleaning.
Note: The shock transmitter insert of two-part transmitter caps can
only be dismantled and the sealing rings can only be removed using
special tools. You should avoid doing this because it could result in
damage to the shock transmitter. It is not necessary for cleaning
purposes.
. Now, from the overhaul kit, take out the new sealing
rings and for the shaft and install them. Observe the O-ring
Guide for this purpose. It is contained in the overhaul kit.