Freeze-dryer Beta 1-8 LSCbasic
Freeze-dryer Beta 2-8 LSCbasic
Version 03/2018, Rev. 1.5 of 09/03/2020 • sb-dc
Translation of the original operating manual
Fig. 44: Dialogue box "Operating hours" (here: refrigeration unit 1.1)
Failure memory (2)
The failure memory stores the most recent messages of the process and
equipment information system. The messages can be viewed in this
dialogue box. The failure memory includes the last 32 messages. If this
number is exceeded, the oldest message will be overwritten.
Use the arrow keys to select the individual messages.
The error message text is always followed by an error code.
Always indicate the error code in the event of enquiries or service
requests!
Fig. 45: Dialogue box "Failure memory"
Vacuum pump oil change (3)
The system monitors the oil change interval of the vacuum pump. The
interval can be adapted to the vacuum pump model and utilisation.
When the end of an oil change interval is reached, a corresponding
message will be displayed.