Calibration Procedure
4-1
4. Calibration Procedure
Before performing this section the HiPot tester must warm up at least for 30 minutes. Take
off the calibration label at the front panel and press the lock switch. After completed the
calibration, please press the lock switch again for opening the hardware data backup
protection circuit to avoid calibration data loss.
The following items are required for calibration.
Voltage Calibration (See section 4.2)
ACV 5KV Offset (0.05kV) ; AC HiPot Voltage Offset calibration (AC Mode)
ACV 5KV Full (4KV) ; AC HiPot Voltage Full calibration (AC Mode)
DCV 6KV Offset (0.05kV) ; DC HiPot Voltage Offset calibration (DC/IR Mode)
DCV 6KV Full (4KV) ; DC HiPot Voltage Full calibration (DC/IR Mode)
IRV 1KV Offset (0.05kV) ; IR Impedance Voltage Offset calibration (DC/IR Mode)
IRV 1KV Full (1kV) ; IR Impedance Voltage Full calibration (DC/IR Mode)
Current Calibration (See section 4.3)
ACA 3mA Offset (0.12mA) ; AC HiPot current 3mA range Offset calibration (AC Mode)
ACA 3mA Full (2.4mA) ; AC HiPot current 3mA range Full calibration (AC Mode)
ACA 20mA Offset (2.4mA) ; AC HiPot current 20mA range Offset calibration (AC
Mode)
ACA 20mA Full (12mA) ; AC HiPot current 20mA range Full calibration (AC Mode)
DCA 3mA Offset (0.12mA) ; DC HiPot current 3mA range Offset calibration (DC/IR
Mode)
DCA 3mA Full (2.4mA) ; DC HiPot current 3mA range Full calibration (DC/IR
Mode)
DCA 5mA Offset (2.4mA) ; DC HiPot current 5mA range Offset calibration (DC/IR
Mode)
DCA 5mA Full (4.8mA) ; DC HiPot current 5mA range Full calibration (DC/IR
Mode)
ARCing Calibration (See section 4.4)
AC ARC 20mA (7mA) ; AC Withstanding Voltage ARCing calibration (AC Mode)
DC ARC 5mA (5mA) ; DC Withstanding Voltage ARCing calibration (DC Mode)
Insulation Resistance Calibration (See section
4.5)
IRR Range1 (1000MΩ) ; IR Resistor range 1 calibration (IR Mode)
IRR Range2 (100MΩ) ; IR Resistor range 2 calibration (IR Mode)
IRR Range3 (10MΩ) ; IR Resistor range 3 calibration (IR Mode)
IRR Range4 (10MΩ) ; IR Resistor range 4 calibration (IR Mode)
Ground Continue Calibration
GC 5Ω Offset (1Ω) ; GC offset calibration
GC 5Ω Full (3Ω) ; GC 5Ω full calibration
LCD Contrast Calibration
CAL. CONTRAST (7) ; LCD contrast calibration

Table of Contents

Other manuals for Chroma 19071

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 19071 and is the answer not in the manual?

Chroma 19071 Specifications

General IconGeneral
BrandChroma
Model19071
CategoryTest Equipment
LanguageEnglish

Related product manuals