Calibration Procedure
4-11
(10) Press [STOP] to stop IRR range 2 calibration.
(11) Press [UP] to change to the next calibration item.
(12) Change the standard dummy load resistor to 10MΩ.
(13) Press [STOP] [START].
(14) Press [INC.] or [DEC.] until shows the standard resistance value, ex. 10MΩ and then
press [ENTER] for saving.
(15) Press [STOP] to stop IRR range 3 calibration.
(16) Press [UP] to change to the next calibration item.
(17) Press [STOP] [START].
(18) Press [INC.] or [DEC.] until shows the standard resistance value, ex. 10MΩ and then
press [ENTER] for saving.
(19) Press [STOP] to stop IRR range 4 calibration.