EasyManua.ls Logo

Chroma 19073 - Ground Continue Calibration

Default Icon
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HIPOT Tester 19071/19072/19073 Users Manual
4-12
4.6 Ground Continue Calibration
4.6.1 Connection Diagram
4.6.2 Description of Ground Continue Calibration
Connect the standard dummy load resistor to the rear panel of HiPot tester which is
between ground continue current output terminal and grounding terminal. Using the
standard 1 and 3 1/8 Watt dummy load resistors.
Prepare a standard 1Ω resistor
(1) Press [UP].
(2) Connect resistor to ground and CONT.CHK OPTION terminal of rear panel.
(3) Press [STOP][START] to read GC resistance value 1.
(4) Press [UP] or [DOWN] until the reading is the actual resistance value.
(5) Press [ENTER] for saving the calibration value.
(6) Press [STOP] to stop GC 5 Offset calibration.
Change the standard resistor to 3.
(1) Press [UP].
(2) Connect resistor to ground and CONT.CHK OPTION terminal of rear panel.
(3) Press [STOP][START] to read GC resistance value 3.
(4) Press [UP] or [DOWN] until the reading is the actual resistance value.

Table of Contents

Other manuals for Chroma 19073

Related product manuals