Page #89 background imageLoading...
Page #89 background image
RS485 Interface (For Link Model of 19073 + RS485)
5-23
Example: Master (0x70): 0xAB 0x01 0x70 0x01 0x7F 0x0F
Slave (0x01): 0xAB 0x70 0x01 0x02 0x7F 0x00 0x0E
Descriptions of example
Inputted setting meanings of HEX are as the following:
0xAB = Header
0x01 = Destination Address
0x70 = Source Address
0x01 = Data Field Length
0x7F = Command Code
0x0F = Checksum
The reading meanings of returning HEX code are as the following:
0xAB = Header
0x70 = Destination Address
0x01 = Source Address
0x02 = Data Field Length
0x7F = Command Code
0x00 = “00”OK, no error (Return Data)
0x0E = Checksum

Table of Contents

Other manuals for Chroma 19073

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 19073 and is the answer not in the manual?

Chroma 19073 Specifications

General IconGeneral
Model19073
CategoryTest Equipment
Output Frequency50Hz/60Hz
Power Rating500VA
AC Test Voltage5 kVAC
DC Test Voltage6kV
IR Test Voltage50 V to 1 kV
InterfaceRS-232, USB
Insulation Resistance1MΩ to 50GΩ
TypeElectrical Safety Tester