EasyManua.ls Logo

Chroma 3302 - How To Set Up〝DEVIATION〞Function

Default Icon
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Automatic Transformer Test System 3300/3302 Users Manual
4-38
4.4.11 How To Set Up
DEVIATION
Function
When users have their sample transformer to be a standard, the sample must has done the test
of each test parameter data before to be a measured standard, then users can use deviation to
error compensation function, the set up procedure as following:
a. According to above method, the screen show testing display after each parameter is set up.
b. Press
DEVIATION
key to ON state on the front panel, the display will be as follow:
c. Make selection of deviation item by tick
& cross
X
key.
d. Press
DEVIATION
key to confirm the function is ON.
SETTING
e. Press EXIT key to enter <SCAN TEST: MEASUREMENT> for auto transformer scanning,
under this state. work
DEVIATION
will become back-lighted shown as figure 5-2a.
(Date show underneath of DEVIATION is what the time users have use this function)
Put the standard transformer on the test fixture and press the key [Start], in first measuring, it
will display standard transformer measured value, and
DEVIATION
back-lighted will
disappear, it means the deviation error compensation is done. In second time you press
[Start] key you will see the value is deviated.

Table of Contents

Related product manuals