EasyManua.ls Logo

Chroma 3302 - Data Output Format for 33 XX

Default Icon
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RS-232C Control Interface
8-5
Rn Read the setting of set n
n = 0
159
Sn Write the setting of set n
n = 0
159
T1 Enable the measurement cycle
T0 Disable the measurement cycle
<Note>Every command has to use RedDrAgOn(ASCII) as the end character code.
8.5 Data Output Format for 33XX
8.5.1 Transformer Output Format
The measurement value of transformer has 509 Bytes and 33XX will output it automatically
after a measurement cycle is completed. The output format is shown in Table 8-4.
Table 8-4 Output Format for Low Voltage Parameter Measurement
Byte Parameter Meaning
0
39 TR1
TR10
Turn ratio and phase degree test. 4 bytes per data
with format in floating point.
40
79 Lx1
Lx10
The inductance quantity. 4 bytes per data with
format in floating point and unit in [H].
80
119 Lk1
Lk10
The leakage inductance. 4 bytes per data with
format in floating point and unit in [H].
120
159 Cx1
Cx10
Pin stray capacitance. 4 bytes per data with format
in floating point and unit in [F].
160
199 DCR1
DCR10
DC resistance. 4 bytes per data with format in
floating point and unit in [
Ω
].
200
239 Q1
Q10
Quality factor. 4 bytes per data with format in
floating point.
240
279 ACR1
ACR10
AC resistance. 4 bytes per data with format in
floating point and unit in [
Ω
].
280
319 Zx1
Zx10
AC absolute impedance. 4 bytes per data with
format in floating point and unit in [
Ω
].
320
359
Reserved.
360
399 PS1
PS10
Pin short-circuit test. 4 bytes per data with format
in floating point.
400
409
Turn ratio and phase
degree judgment
0: GO, 1: NG
410
419
Inductance judgment 0: GO, 1: NG
420
429
Leakage inductance
judgment
0: GO, 1: NG
430
439
Pin stray capacitance
judgment
0: GO, 1: NG
440
449
DC resistance judgment
0: GO, 1: NG
450
459
Quality factor judgment
0: GO, 1: NG

Table of Contents