Programmable AC Source 6430/6420/6415 User‘s Manual
Return Parameters: 0 | 1
Query Example: VOLT:RANG:AUTO? It returns the condition of AUTO range mode.
STATus:PRESet
Description: This command sets the PTR, NTR and Enable Registers of Questionable and
Operation status as follows: all defined bits of PTR Registers to 1, all bits of NTR
and Enable Registers to 0.
Syntax: STATus:PRESet
Parameters: None
Example: STAT:PRES It presets the PTR, NTR and Enable Registers of Questionable
status.
STATus:OPERation?
Description: This query returns the content of the Operation Event Register. The Operation
Event Register is cleared when read or by the *CLS common command.
Query Syntax: STATus:OPERation[:EVENt]?
Parameters: None
Return Parameters: <NR1>
STATus:OPERation:CONDition?
Description: This query returns the content of Operation Condition Register. The Condition
Register reflects the TRUE or FALSE state of Operation status.
Query Syntax: STATus:OPERation:CONDition?
Parameters: None
Return Parameters: <NR1>
STATus:OPERation:ENABle
Description: This command sets the enable mask which allows true conditions in the Operation
Event Register to be reported in the Operation Status (OPER) Summary-Message
of the Status Byte Register.
Syntax: STATus:OPERation:ENABle <NRf>
Parameters: 0 to 32767
Query Syntax: STATus:OPERation:ENABle?
Return Parameters: <NR1>
6-14