MAINTENANCE WITH Q & A
80
WATER QUALITY
Purified water is recommended for use with our humidity systems. Water
should be provided within 0.05 to 2M (.25 to 10ppm) DI and <2mg/L of free
chlorine, Distilled water or Reverse Osmosis (RO) water outside of these limits
may cause either corrosion or scaling. Tap water may be used with our optional
Demineralizer Filtration System. A Recirculation Humidity Water Supply System
is also available. Inlet water pressure should not exceed 10 PSI for use with an
atomizer or 25 PSI for use with a boiler. For optimal use of the atomizer and
boiler systems, inlet water pressure should not fall below 10 PSI (69 kPa) for
both boilers and atomizers. A Water Pressure Regulator must be used.
Water specification 0.05 to
2 Mega OHMS in addition
to <2mg/L of free chlorine.
CAUTION
Maximum water inlet
pressure is 10 PSI (69kPa)
for an atomizer and 25 PSI
(172 kPa) for a boiler.
Exceeding this pressure
may cause catastrophic
failure of the filter housing.
For optimal use of the
atomizer and boiler
systems, inlet water
pressure should not fall
below 10 PSI (69 kPa) for
both boilers and atomizers.
WARNING
Failure to maintain cartridge
may result in chloride
corrosion of stainless steel
interior surfaces, which is
not covered under warranty.
WARNING