Installation area requirements
12
4.
Installation area requirements
4.1.
Environmental requirements
m
Place the generators FAR FROM SOURCES OF HEAT, PROTECTED
AGAINST RAIN AND WIND, and SHOULD NOT BE EXPOSED TO THE
SUN.
e
Failure to observe the precautions preclude the guarantee.
4.2.
Environmental ventilation
The environment where the generator is installed must be good ventilated to
prevent the risk of asphyxia and decreased the performance of the generator.
4.2.1.
Oxygen content in the ambient
m
If the generator and the point-of-use are located in different rooms,
the following minimum ventilation air flow must be provided to main-
tain acceptable oxygen content in the ambient. Don’t seal the rooms.
4.2.2.
Temperature
e
This paragraph is able only for these models: NiGen LCMS 40-1 and NiGen HF-1
m
Claind raccomanded to not go over the maximum environment tempe-
rature of 30°C for guarantee correct operations of the generator. If
this value is exceeded, in relation to the use of the generator, could ap-
pare the warning of maximum temperature of the compressor (E03)
(par.9.4.).
Are requires two openings of at least 0,5m
2
to create a flow of air suffucient to
ensure air exchange.
It is also recommended to avoid danger of asphyxia.
Installation site enclosed, indoors
Relative humidity up to 80%, no condensation
Use temperature from 5°C to 35°C from 40°F to 95°F
Protection rating IP20