Nitrogen Generator NiGen HF - NiGen LCMS
19
6.
Disassembly and transport
6.1.
Disassembly
Stop the generator
(par 7.6.);
wait for depressurization (about a minute);
switch off the generator
(par. 7.7.);
disconnect the electric power cable;
only for NiGen HF-1 and NiGen LCMS 40-1
, depressurize the air tank, removing
the air outlet plug
(par. 3.3.2.)
and acting on the air outlet pressure regulator as
described at
par. 7.9.2.
Plug again the air outlet;
carefully detach the pneumatic connections.
6.2.
Transport
e
Remember that during transport the generator must always remain in a vertical
position.
If conserved, use the original packaging; otherwise use a pallet of adequate di-
mensions to hold the generator, affixing instructions in visible locations, such as:
THIS WAY UP, KEEP IN VERTICAL POSITION.