Troubleshooting
36
9.4.
List of errors
error identification cause resolution
E-00 type error generator not identified
assign a number to the
generator (
see Stack user’s
manual
)
E-01 low inlet pressure
inlet air pressure below mini-
mum operative value
(only for:
NiGen LCMS 40-0; NiGen LCMS
100; NiGen HF-0)
increase inlet air pressure
E-02 high inlet pressure
inlet air pressure above maxi-
mum acceptable value
(only
for: NiGen LCMS 40-0; NiGen
LCMS 100; NiGen HF-0)
decrease inlet air pressure
E-03 compressor tempe-
rature high
compressor box temperature
above threshold
(only for:
NiGen LCMS 40-1; NiGen HF-1)
contact service
E-04 air tank pressure
high
air tank pressure above maxi-
mum threshold
(only for:
NiGen LCMS 40-1; NiGen HF-1)
contact service
E-05 air tank pressure
low
air tank pressure below mini-
mum threshold
(only for:
NiGen LCMS 40-1; NiGen HF-1)
contact service
E-06 on chip error control board can not read data contact service