EasyManua.ls Logo

Claind NiGen LCMS - Maintenance Schedule

Claind NiGen LCMS
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Nitrogen Generator NiGen HF - NiGen LCMS
31
8.3.
Maintenance schedule
e
If a correct maintenance schedule is not followed, the performance of the gene-
rator may no longer be ensured and it may lead to permanent damage to the
generator. The following table specifies the frequency of the recommended
maintenance
operations, express in “working hours” of the compressor.
e
Lack of maintenance will void the warranty.
m
maintenance operations must be performed by authorized personnel
8.3.1.
NiGen HF-1 and NiGen LCMS 40-1
Components use versus maintenance plan
Component part
number
Description 2000 h 4000 h 6000 8000
R0539021 Intake air filter cartridge 1 1 1 1
R0539022 Intake air filter cap 1 1 1 1
R062999001
Silicone disk NRV
16X6mm
1 1
R0539023 Motor capacitor 1
R063399
Filter cartridge Syntesi 5
micron
1 1 1 1
R063400
Filter cartridge Syntesi
20 micron
1 1 1 1
R0539024
Dumping feet kit (3 pie-
ces)
1 1
R061084 CMS silencer 3/8 1
R700.M4.7060 Compressed air cooler 1
R0539025
Maintenance compres-
sor kit
R0539027
Maintenance compres-
sor kit
Maintenance kit to be used --> 422.93.0025 422.93.0026