Software ver. CVPCA30E_1 18
Door Open – Alarm Test
Insert a BodySet® in the pump and close the door. Press the door until a click is
heard.
NOTE: For instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press
to confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
Start infusion as instructed in the "Select protocol" section.
Door Open alarm will be activated at once.
Close the door and press .
Battery Test
Insert a BodySet® in the pump and close the door. Press the door until a click is
heard.
NOTE: For instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
Take the pump out of the charger.
Press until the Self-Test screen appears. If air sensor is off, press
to confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
Start infusion as instructed in the "Select protocol" section.