Software ver. CVPCA30E_1 18
Door Open Alarm Test
1.
Insert a BodySet® in the pump and close the door. Press the door until a click is
heard.
NOTE: For instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
2.
Press until the Self-Test screen appears. If air sensor is off, press
to confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
3.
Start infusion as instructed in the "Select protocol" section.
4.
Open the pump door.
5.
Door Open alarm will be activated at once.
6.
Close the door and press .
Battery Test
1.
Insert a BodySet® in the pump and close the door. Press the door until a click is
heard.
NOTE: For instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
2.
Take the pump out of the charger.
3.
Press until the Self-Test screen appears. If air sensor is off, press
to confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
4.
Start infusion as instructed in the "Select protocol" section.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 575 Color Vision and is the answer not in the manual?

CME BodyGuard 575 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 575 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals