Software ver. CVPCA30E_1 35
12.
Press_ to confirm.
13.
Scroll to select Protocol required then press .
14.
Review screen is displayed showing infusion parameters for the protocol
you selected. Check infusion rate, bolus volume, lockout time and 4 hour
limit are correct before proceeding.
NOTE: When MediGuard® is enabled and the max. rate allowable for the
current patients weight has been exceeded than the pump wil alert
"MediGuard® Limit".
NOTE: When selecting a standard pre-programmed protocol ALWAYS
check the review screen carefully to ensure the parameters match the
prescription.
15.
Press to commence infusion.
Titrating Basal Rate During Infusion
To change Rate during the infusion, enter the new rate using the keypad and confirm by
pressing the key. Confirm again the new rate by entering the Relevant Code. The
Rate can be increased up to the “Max Titration Rate” programmed into the specific
infusion. The pump will beep to confirm acceptance of the change and the new rate will
be displayed.
NOTE: If the desired rate is above the protocol maximum range the pump will not
accept the entry and will display a message notifying the maximum rate
allowed.
Patient Activated Bolus
The BodyGuard® infusion pump can be programmed to allow the patient to request
boluses of a specified volume within predetermined time intervals. The patient can
request the bolus by using the following method:

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 575 Color Vision and is the answer not in the manual?

CME BodyGuard 575 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 575 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals