Software ver. CVPCA30E_1 45
If using an ambulatory administration set without a drip chamber, priming
the line might be needed. Disconnect the set from patient before priming.
Eliminating Air During Operation
When air is detected in the line, the pump gives an audible alarm, the screen
shows AIR IN LINE, and the LED indicator turns from green to red.
To eliminate air in the line during pump operation, follow these steps: -
1. Press _ to cancel the alarm. Press and hold down for 3
seconds again to return you to the Resume/Menu option screen.
2. Check and resolve the cause of air detected in the line.
3. Press _ _ again and enter the relevant Access Code.
4. Disconnect the line from the patient taking care to preserve sterility.
5. Select ‘Prime’ and press .____
__ .
WARNING: Ensure the line is disconnected from the
patient before proceeding to the next step. Failure to do so
could result in the patient receiving a large air bolus.
6. Press within 10 seconds to commence priming, otherwise the
display will back 1 step automatically. You may cease priming at any time
by pressing .
7. Repeat steps 4 & 5 until all air is eliminated from the line and fluid is seen
at the distal end of the set.
8. Press to return to the Resume/Menu option screen.
9. Press _ to resume the current infusion.
10. Check the Infusion Summary if screen which shows: bag volume, volume
left and infused, is correct then press again to restart delivery.
Down Occlusion Alarm
After Down Occlusion Alarm press .