6-Specifications
Software ver. CVPCA30E_1 51
6 Specifications
Pumping Mechanism:
Piston Pump
Flow Rate:
0 to 100 ml/hr in 0.1ml increments
Priming Rate:
300ml/h to 425ml/h (default 400ml/h)
Total Infused Volume:
0.1 to 1,000 ml.
Total Time Setting:
100 hours
Accuracy:
±5%
KVO rate:
0.1 to 5 ml/hr
Air Sensor:
Ultrasonic, adjustable air bubble size
Maximum Pressure:
10 psi Adjustable (250mmHg 520mmHg
(5psi 10psi )).
Power Supply:
110-240 VAC, 50/60 Hz. 0.3A max.
Battery:
Rechargeable Li-Ion 7.4V, 1800mAh. (Typical)
Battery Operation
17 hours at 100ml/hr (Rechargeable)
Battery Charging:
Automatic when clicked into the Charger that is connected to an
AC power source. Six hours needed to charge a fully-depleted
battery.
Alarms:
When a problem is detected, the BodyGuard® displays the
following alarms:
Air in line
Down Occlusion
Pump Unattended
End of infusion
Low Battery
End Battery
Door Open
Error (followed by error code)
Lock Mode
Lock Out Mode
Missing Key

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 575 Color Vision and is the answer not in the manual?

CME BodyGuard 575 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 575 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals