Software ver. CVPCA30E_1 22
Dedicated BodyGuard® Administration Sets
The BodyGuard® Infusion System should be operated only with BodySet® dedicated
Infusion sets.
Warning: Do not operate the BodyGuard® Infusion System with any
administration set, other than the BodySet®/MicroSet®. The
BodySet®/MicroSet® Infusion Sets are equipped with a free flow
protection valve, to prevent a gravity free flow hazard. Using a
different set may essentially change the delivering rate and expose
the patient to a free flow rate, which can cause serious injury or
death.
Warning: Using the pump with air in line detector off, may cause an embolism
resulting death or paralysis. Switching the air detector off is not
recommended and should be used only when absolutely
necessary. Please contact your technician in order to enable the air
sensor if disabled with no need.
The BodyValve
The Anti-Free Flow valve, BodyValve, enhances pump functioning by:
Preventing free-flow in the event the set is detached from the pump.
Preventing back-flow (reflux) in the event several infusion pumps are connected
simultaneously to the same patient.
Preventing free-flow in the event of a mechanical malfunction.
Preventing pump operation if the set has been loaded incorrectly.
Warning: Replace the administration set every 72 hours to lessen the
incidence of bacteria formation.
Warning: Disposables must be compatible with the medicine delivered.
Warning: Patient or operator injury may result if package is opened or
damaged, or if damaged components are used. Visually inspect
contents and package before each use.
Warning: Air embolism can cause death or serious injury to the patient. Do not
connect a patient to the BodyGuard® Infusion Pump until all trapped
air has been cleared from the fluid path. Carefully read the
instructions for loading. Puncturing set components may cause air
embolism.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 575 Color Vision and is the answer not in the manual?

CME BodyGuard 575 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 575 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals