Software ver. CVPCA30E_1 21
Keypad Lock
To Lock Keypad
Press and hold until the entire graph is
black and a beep is heard. The beep indicates that
the locking program is turned ON.
NOTE: The keys are locked in memory until they are unlocked or until
entering Change setup mode.
To Unlock Keypad
Press and hold until the entire graph is
clear and a beep is heard. The beep indicates that
the locking program is turned OFF.
NOTE: To prevent unauthorized individuals accessing the main menu a technique
is required to return to the menu code screen when the pump is in STOP mode.
To return to the menu during infusion, press to stop delivery and enter
the STOP state.
Press and hold until the ‘Resume or Menu’ choice prompt appears.
NOTE: When keypad lock is on, entering the STOP state & holding down
will not return you to the main menu. Press to restart the
infusion, press and hold to remove keypad lock, then press and
hold until you are prompted for the level 1 access code to return to
the main menu.