6-Specifications
Software ver. CVPCA30E_1 52
Pump unattended
NOTE: When this symbol is seen, consult accompanying documents.
BodyGuard®
Dimensions:
112 x 89 x 32mm. (L x W x H).
Classification
Type CF Equipment (degree of protection against electrical
shock)
Housing:
PCABS (fire retardant)
Weight
280 grams without battery
390 grams with battery
Electrical Safety
Complies with: EN 60601-1 (Medical Electrical Equipment
Safety), IEC 60601-2-24 (Infusion pumps and controllers), IEC
60601-1-4 (Programmable Electrical Medical System)
Standards
Manufactured in accordance to ISO 9001:2008 and ISO
13485:2003 . CE marked (In accordance with the Medical
Devices Directive 93/42/EEC)
EMC
The BodyGuard® Infusion System is designed to be in
compliance with EN 60601-1 (safety) and IEC 601-1-2 (EMC).
Environmental
Specifications
Non Operating Conditions (Transportation and Storage):
Temperature: -25
o
C to 50
o
C (-13
o
F to + 122
o
F)
Humidity: 5 % to 100% R.H., non-condensing
Air pressure: 48kPa to 110kPa
Operating Conditions:
(The system may not meet all performance specifications if
operated outside of the following conditions)
Temperature: +18
o
C to +45
o
C (+59
o
F to + 113
o
F)
Humidity: 20 % to 90% R.H. at +40
o
C, non-condensing
Air pressure: 70kPa to 110kPa
Administration set
Dedicated administration sets with anti-siphon valve.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 575 Color Vision and is the answer not in the manual?

CME BodyGuard 575 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 575 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals