Software ver. CVPCA30E_1 50
Fatal internal error has occurred
Restart the pump. If the alarm does
not stop pack the pump properly, in
its original packaging, and send it for
service.
Program ends,
pump will turn to
KVO mode
Current infusion program has
been completed.
Press STOP/NO to restart a new
program or turn the pump OFF.
Administration set loaded
incorrectly (the key was not
located in its place correctly).
A set different than the
BodySet®/ MicroSet® was
loaded in the pump.
Reload the set into the pump as per
the instructions.
Change the set to a BodySet®/
MicroSet® and load it according to
instructions.
Setting cannot be
changed
Lock Out mode is turned on.
No action required. If changing
program is required, unlock the pump
and refer to lock section for further
instructions.
Setting keys do
not function
Lock Out mode is turned on.
No action required. If changing
parameters is required, unlock the
keys and refer to lock section for
further instructions.