Software ver. CVPCA30E_1 49
Troubleshooting
Infusion stops
and an alarm is
activated
Air is present in administration
Set.
The roller or clip on the
administration set is closed
upstream of the pump.
The line was not loaded correctly.
Disconnect line from patient and
press STOP/NO key to move to
Pump stopped screen. A long press
(3sec) on the STOP/NO key
followed by a normal press on
STOP/NO key will return to the
priming screen as described on page
25 (through level one code screen).
Re-load the administration line.
Infusion stops
and an alarm is
activated
The roller or clip on the
administration set is closed
downstream of the pump.
Administration set is kinked.
Cannula is blocked.
Administration set loaded
incorrectly.
NOTE: Operator may change
pressure setting.
Open clamp/clip.
Straighten the set.
Change the Cannula.
Re-load the administration line.
Refer to “Setting Down Occlusion”
section.
Two minutes has elapsed
without a button press during
programming.
Press START/OK to resume.
Infusion stops
and an alarm is
activated
The door of the pump was not
closed prior to operation
The door has accidentally
opened during operation.
Close the door of the pump.
Send the pump for service.
Pump will only
run for another
30 minutes if it is
not connected to
mains.
30 Minutes of battery life are
remaining.
Place the pump in charger and
connect to mains.
Pump operation
stops. The pump
cannot be used
before being
connected to
mains.
Place the pump in charger and
connect to mains. Wait 2 minutes
before operating.