Statox 503 Control Module
Page 20 of 46 Issue 10 / 2018
Calibration
In this menu you can perform a calibration of sensor heads type Statox 501 HRC, ARE, LCIR, MCIR, CO
2
and
PID detecting combustible gases, carbon dioxide (CO
2
) and volatile organic compounds (VOC). The test gas
concentration must be within the permitted range, see program list.
CALIB.
ROUTINE
ZERO
ADJUSTM.
000
APPLY
SPAN GAS
GAS IS
ON
R
R
CALIB.
FINISHED
GAS CONC
(value) (unit)
FACTOR
x.yy
DONE
R
CALIB.
FAILED
GAS
(value) (unit)
R
R
GAS CONC
ERROR
R
ERROR 6
ERROR 7
ERROR 8
APPLY
SPAN GAS
R
Figure 14: Calibration menu diagram
Change program
Start zeroing
Flashing during zero procedure
Zero stable
Flashing, span gas requested
Span gas detected
Plateau found. Shut off span gas
Enter span gas concentration
Enter response factor. Correctable with
“R”. Confirm with “ENTER”.
Displayed for 2 s , calibration successful
Actual concentration
Measuring Mode