Statox 503 Control Module
Page 24 of 46 Issue 10 / 2018
1. Enter the menu „SENSOR POWER“ by pushing .
2. Display „SENSOR ON“.
Up to step 7 the automatic time-out is not active.
3. Change the status of the sensor head power supply with ◄ or ►.
4. Confirm with Enter.
5. „SENSOR OFF“ flashing tells you that the sensor head power supply is off.
6. Push to switch the power supply on again.
7. „SENSOR ON“ tells you that the sensor head power supply is on.
8. Push Reset to return to the main menu and again to return to measuring mode.
Programming the alarm relays
In this menu you can set parameters of the alarm relays A1 and A2:
1. Alarm thresholds. A1 must be lower than A2.
2. High or low alarm
3. Latching (HOLD) or non-latching (AUTO RESET)
4. Coil active (ACTIVE) or non-active (PASSIVE)
The ex – works settings of the alarms are listed in the program list, which comes with each control module,
or is accessible in the internet: www.compur.com
If an alarm threshold is exceeded, the LEDs A1 and A2 are on, and the relays switch. Alarms can be reset,
provided the reason for the alarm has disappeared. If the alarms have been programmed to be latching, you
have to push reset button or use the external reset to delete the alarm message. For additional information
see chapter 4.5.1.
!