134
TECHNICAL
SPECIFICATIONS
G
5.2 Controller
I.E.C. Equipment Classification: Patient Applied, continuous operation with intermittent loading
as Class 1, Type B per IEC 601-1, UL 2601-1 and CSA 601-1.
The system was tested and complies with IEC 601-1-2 as follows: Operation is subject to the
following two conditions: (1) The system may not cause harmful interference and (2) The system
must accept any interference received, including interference that might cause undesirable operation.
The system is considered fail-safe, and can recover from error mode by power-down restart or by a
method described to the user via the system display.
5.2.1 Advantage Controller
Typical Operating Requirements: D3000
D3000I D3000J
Input Voltage: 115 VAC 230 VAC 100 VAC
Frequency: 50/60 Hz 50/60 Hz 50/60 Hz
Power Consumption: 500 VA 500 VA 500 VA
Line Frequency Leakage: Less than 100 ma
Fuses: Reference the specification label located on the back of the
controller.
Output: 3 phase motor drive and control circuits, 325 VA
Port 1 Output: 180 VA
Port 2 Output: 180 VA
Port 3 Output: 325 VA
Total Port Output: 325 VA max. (With 2 handpieces simultaneously connected)
Dimensions: 17.0” (W) x 5.0” (H) x 13.75” (D)
43.2 cm (W) x 12.7 cm (H) x 35.0 cm (D)
Weight: 19.5 lbs. (8.86 kg)
Pump Flow Rate: 0 - 150 mL/minute