70
E9000
HANDPIECES
B
E9415A Tubing Set with Integrated
Irrigation Tip:
NOTE: The integrated irrigation tip of this
tubing set will attach to all the associated bur
guards for the CoolFlex High Speed Drill.
1. SCRUB - Peel the Discharge Line tubing,
marked “TO WASTE CONTAINER” away
from the rest of the tubing and pass to the
Circulator.
2. CIRCULATOR - Attach this line to the
discharge bucket using the attached waste
container adapter.
3. SCRUB - Prior to attaching the irrigation
tip to the bur guard, bend the irrigation tip
to the desired position and lightly squeeze
the black clips and slide them to the desired
location.
NOTE: When using the medium bur guard
(E9011-1) remove the distal clip from the
tubing).
4. SCRUB - Snap the clips onto the bur guard.
Note: The rear clip has an integrated rib
that will snap into the rear groove of the
bur guard.
5. SCRUB - Snap the handpiece cord clips
onto the handpiece cable to secure the
tubing.
6. CIRCULATOR - To adjust fluid flow to
the operative site between 40% and 100%,
use the controller Flow Increase or
Decrease buttons.
7. SCRUB - To adjust fluid flow to the
operative site below 40%, use the roller
clamp.