9
The OLED displays “BRIGHTNESS”
via a bar graph in increments of 5%.
The BRIGHTNESS Up Arrow
increases brightness in increments
of 5%.
The BRIGHTNESS Down Arrow
decreases brightness in increments
of 5%.
The STAND-BY button returns the
light source to STAND-BY mode. The
POWER button turns the light source
off.
4. Light Source ON, Auto Shutter Control:
To enable Auto Shutter control, the
ethernet cable must be connected
between an IM4000 series ConMed
Linvatec camera and the light source.
The IM4000 series camera head button
can enable or disable the STAND-BY
mode.
4a When STAND-BY mode is enabled:
OLED displays “STAND-BY” and
remaining lamp life hours.
The BRIGHTNESS buttons are
inactive. The POWER button will
turn the light source OFF. The
STAND-BY button returns the light
source to Auto Shutter mode.
4b. When STAND-BY mode is disabled:
The BRIGHTNESS buttons are
controlled by the IM4000 series
camera to achieve the desired picture
brightness.
The STAND-BY button returns the
light source to STAND-BY mode.
The POWER button will turn the
light source OFF.
5. Lamp Life:
When the light source is in the ON
condition and the STAND-BY mode is
enabled, the OLED displays lamp life
hours remaining, beginning at 500
hours and counting down to zero in
increments of one hour.
When lamp life is 10 hours or less, the
OLED displays a blinking message to
indicate nearing end of life or at end of
life.
6. Lamp Replacement:
For installation instructions see
“3.3
Lamp Replacement” on page 10
.
Replace only with ConMed Linvatec
Lamps. Use of unauthorized lamps will
void certifications and/or warranties
and may affect safety and performance.
7. Lamp Door Open:
Lamp door must be closed for unit to
operate.
8. Lamp Saver:
If light source is in STAND-BY mode,
continuously for one hour, the light
source lamp will turn off. When
STAND-BY mode is disabled, the
lamp will turn on and the light source
will return to normal operation.