3.5 Troubleshooting
Symptom Possible Cause Corrective Action
Lamp will not ignite.
Power cord is not plugged in or
is faulty.
Power is not turned on.
Lamp access door open.
Lamp module is not fully
engaged.
Lamp is expired.
System overheats or lamp has
expired.
Plug power cord into receptacle
and/or Light Source.
Replace power cord.
Place POWER button in the on
position. LED will illuminate
green when unit is on.
The Light Source will not work
with the lamp access door open.
Assure door is completely closed
and latched.
Verify lamp assembly is
completely engaged on
receptacle pins within Light
Source.
Replace lamp (see
“3.3 Lamp
Replacement” on page 10
).
Turn power off. Ensure intake
vents are not blocked. Allow
system to cool for five minutes.
Turn power back on. If lamp
does not ignite, replace lamp
assembly.
Use of unauthorized lamp
module. Use only ConMed
authorized replacement lamp
modules.