14
Low light output.
Brightness is set to minimum or
too low of an output position.
Fiber optic light guide surface(s)
is contaminated or damaged.
Lamp life exceeds warranted
life.
Light guide is not engaged
properly.
Excessive number of light guide
fibers are broken.
Increase lamp brightness.
Replace light guide.
Check lamp life indicated on
OLED. Replace lamp.
Check to ensure light guide is
fully inserted into LGR.
Inspect light guide cables. If
more than 20% of fibers are
broken (darkened), replace light
guide.
Lamp is unstable
(flashes).
Lamp life exceeds warranted
life.
Lamp problem.
Check lamp life indicated by
OLED. Replace lamp.
Replace lamp.
Symptom Possible Cause Corrective Action