17
Power Cord Requirements:
100/120 Volt (Cat. No. C7104)
Use only a listed (UL, CSA) detachable power cord manufactured to the following specifications.
• Plug End
• NEMA 5-15P hospital grade, 15 amps, 125V
• Receptacle End
• IEC 320/CEE-22, 6 amps, 250V/15 amps, 125V
• Cord
• UL style SJT, 14 AWG, 3 conductor
220/240 Volt (Cat. No. C7105)
• Plug End
• Molded straight PVC plug with double grounding system
• DIN 49441, CEE 7/U11, 10/16A, 250V
• CEBEC, DEMKO, KEMA, NEMKO, OVE, SEMKO, VDE, UTE, FEMKO
• Receptacle End
• Molded straight PVC plug
• DIN 49457, CEE 22/V, 10A, 250V
• VDE, D, N, S, SEV, OVE, KEMA
• Cord
• PVC, 7.2mm diameter
• 10A, 250V
• Conductors: 3 x 1 mm
2
• Conductor Colors - brown, blue, green/yellow stripe
NOTES:
1. If the product is connected to a 240V system in the United States, then it must be connected
to a center tap system.
2. This light source contains electronic components and may require special handling for
end-of-life disposal. Refer, for example, to Directive 2002/96/EC (WEEE guidelines) for
disposal in the European Union or other local guidelines regarding disposal of electronic
components.