21
Interference may occur in the
vicinity of equipment marked with
the following symbol.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and/or people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the LS7700 Light Source is used exceeds
the applicable RF compliance level above, the LS7700 Light Source should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the LS7700 Light Source.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 2: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ConMed Linvatec LS7700 Light Source is intended for use in the electromagnetic environment
specified below. The customer or the user of this LS7700 Light Source should assure that it is used in
such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environment
guidance