Cosman G4 Radiofrequency Generator Service Manual (SPI 11291 Rev A)
Section 5
5.3 Testing the G4 Unit
The ability of the generator to produce RF output and to measure Temperature and Impedance can be
tested by attaching one or more Test Plugs (RFG-TP) to active electrode jacks and performing a procedure.
When a Test Plug is attached to an active electrode jack and the output is off, check that the displayed
Impedance is approximately 100 and the displayed Temperature is approximately room Temperature.