EasyManua.ls Logo

Cosman G4 - Page 33

Cosman G4
115 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cosman G4 Radiofrequency Generator Service Manual (SPI 11291 Rev A)
8.3.1 Additional Example Screens
The Graphing Pulsed RF screen will appear differently depending on the current settings:
Electrode Setup settings (see Section 7.1)
E-dose setting: Vary Voltage, Vary Width, or Vary Rate (see Section 8.3.2)
Pulse Width Reading:
Displayed in units of
milliseconds only if E-dose =
Vary Width. The Pulse Ra
is fixed at the value shown
in the Settings Panel.
te
Electrode Panel (Black area):
Displays all readings for the
indicated bipolar electrode pair,
e.g. “Electrode 1+2”.
E-dose = Vary Rate
One Electrode
E-dose = Vary Voltage
One Bipolar Pair
Temperature Readings:
One Temperature reading
from each electrode in
bipolar pair. For instance,
“E1” indicates a reading
from Electrode 1.
Bipolar readings: Voltage,
Current, and Impedance
between the two electrodes in
the bi
p
olar
p
ai
r
Settings Panel: If E-dose =
Vary Width, the Pulse Width
and Pulse Rate are fixed at
the values shown here.
E-dose = Vary Width
F
ou
r El
ect
r
odes
Pulse Rate Reading:
Displayed in units of
Hertz only if E-dose =
V
ary Rate. The Pulse
Width is fixed at the
val
Section 8

Table of Contents