Cosman G4 Radiofrequency Generator Service Manual (SPI 11291 Rev A)
Section 11
Problem Possible Cause
output shuts down prematurely
Temperature above 100 °C.
Break in electrode or electrode cables.
May be intermittent problem.
Set Time setting too low
Maximum RF Current or Power limit
exceeded
Electrodes attached to generator do not
match Electrode Setup
Temperature does not reach Set Temp or
Max Temp setting
Required total RF Current or Power limit
exceeds allowed limit
Total Impedance too low
Too many electrodes energized
simultaneously
Set Temp or Max Temp too high
Temperature can only be controlled on
one electrode in a Bipolar Pair
Ground pad used when RF Type=Bipolar