Cosman G4 Radiofrequency Generator Service Manual (SPI 11291 Rev A)
Section 0
THERMAL & PULSED RF OUTPUT (continued)
Volts: 0 to 70 volts RMS into 100 load during Thermal RF output.
0 to 100 volts peak into 100 during Pulsed RF output.
Resolution: 1 volt RMS during Thermal RF output
1 volt peak during Pulsed RF output
Accuracy: ± 10% of full scale
Current: 0 to 700 mA RMS
Resolution: 1 mA RMS during Thermal RF output
1 mA peak during Pulsed RF output
Accuracy: ± 10% of full scale
PULSED RF SETTINGS:
Repetition Rate: 1, 2, 3, 4, 5, 6, 7, 8 PPS (Hz)
Burst Duration: 0.5, 1, 5, 10, 20, 30 ms
TEMPERATURE MONITOR
Range: 0 – 110°C
Resolution: 1 °C
Accuracy: ± 3 °C between 40 and 90 °C
± 4 °C between 10 and 40 °C, and between 90 and 110 °C
GENERAL
Size: W 14.25” (36.2 cm), H 9.5” (24.1 cm), D 12.15” (30.9 cm)
Weight: 22 lbs (10 kg)