50
“E1-BAD TXT EPROM”; A bad scale may cause this, also.
“E2-SCALE ADC”; Connection to the scale electronics lost.
“E3-SCALE EPROM”; Scale not calibrated?
“E4-CANNOT CAL; Scale cannot be calibrated or bad calibration.
“E5-MBV FAULT: X”; Where X is a number from 1 to 5 indicating which MBV has failed.
“E6-AMB. THRM”; Ambient thermistor in the “head” is bad.
“E7-TNK THRM”; Thermistor attached to the oil sep. is bad.
“E8-PT TRANS”; Pressure transducer is bad.
“E9-WEAK BATT”; Should never appear.
“E10-BAD FILTERTIME”; Filter code problems.
“E11-DEFAULTS NOT SET”; What it says…
“E12-BAD GRAPH MEM”; Problems with the graphics like Chinese and Russian.
“E13-BAD PTR. TO INFO MEM”; Major software corruption, fatal error.
“E15 FLOW SWITCH ERROR” Air flow switch is open.
“UNKNOWN ERROR”; Cause of problem cannot be identified
MX ERROR CODES
Part Number Description
FX3030X1 FX Series Maintenance Kit—Filter, Vacuum Pump oil and Coupler O-rings
QCL134 Auto Coupler Brass Low Side 14mm
QCH134 Auto Coupler Brass High Side 14mm
AR2788X14A R134a Refill Hose Adaptor 1/2 ACME
AR2788X14B R134a Refill Hose Adaptor 1/4 SAE
CRX430T 30lb 400psi Recovery Tank
CRX400T 50lb 400psi Recovery Tank
481500 / 481500YF Spotgun™ Oil Injection Kit
REPLACEMENT PARTS
“00001” Error #1: The air or gas readings were unstable.
Solution: Move the unit away from sources of EMF or RFI
such as radio transmitters and arc welders.
“00002” Error #2: The air or gas readings were excessively high.
Solution: Move the unit away from sources of EMF or RFI
such as radio transmitters and arc welders.
“00003” Error #3: The air calibration resulted in a low output.
Solution: Prevent refrigerant from flowing into the unit
through the sample inlet during air calibration.
Solution: Allow any refrigerant in the atmosphere to
dissipate before performing air calibration.
Solution: Verify that the air intake and the exhaust are
not obstructed
Solution: Verify that the white filter is correctly plugged
into the rubber grommets.
“00004” Error #4: The unit is beyond the operating
temperature range
Solution: Move the unit to an area where the ambient
temperature is within the specified operating range.
“00005” Error #5: The refrigerant sampled has an excessively
large amount of air or there was little or no sample flow due
to a closed valve or plugged sample filter. This is the code
to prompt the user to change the brass filter. This should be
considered more as a prompt than an actual error.
Solution: Verify the coupler valve is open.
Solution: Verify the sample filter is not plugged with debris
or oil.
Solution: Replace brass sample filter
Solution: Verify that the white filter is correctly plugged into
the rubber grommets.
“00006” Error #6: The air sensor has expired and must be
replaced before the analyzer can be used.
“00007” Error #7: The pressure read by the sensor has been
determined to be too high.
Solution: Verify that the exhaust is not obstructed.