MNL-00007 [3] Proprietary and Confidential Page 21 of 63
18.1 The Backup Time
An important safety feature of Daavlin’s Integrated Dosimetry System is the use of a backup
safety time. While the dose is entered in Joules or milliJoules, a backup time must also be
entered as a safety check. The treatment will stop when the dose has been delivered, OR, when
the backup time has elapsed, whichever comes first. At the end of the treatment, the dose
delivered and the time elapsed are alternately displayed so that you will know which limit ended
the treatment.
When preparing for a treatment, the back-up time is entered first. To determine an appropriate
back-up time, it is necessary to know the power output of your device.
18.2 Measuring the Power Output of Your 1 Series
When measuring the power output of your 1 Series, be sure to select the appropriate treatment
distance, “0 in” or “9in”, based on how you will be using the device. If you will be doing
treatments at both distances then you will need to do two power output measurements, one at “0
in” and one at “9 in”.
Note: The power output values of your device, as measured at the factory, are shown on the
Power Output Certificate included with your manual.
Note: It is not necessary to take a power output reading before each treatment. We suggest once
every couple of weeks or earlier if the backup time consistently ends your treatment before the
full dose has been delivered.
1. When just a decimal point is displayed, the device is “asleep“(energy saving feature).
Press MODE or any key to “wake up” the device.
2. The word “CODE” will appear. Enter in your key code. Unless you have changed it, the
factory preset key code will be 0007.
3. Once the safety key code, if activated, has been entered, you will be prompted to select
your desired treatment distance.
4. Press the ▲ key to toggle between “0 in” and “9 in”. Press the ENTER key to lock in
your desired treatment distance.
5. Once you have selected your desired treatment distance press the ▲ key five times. The
screen will read “00:05”.
6. Press the ◄ key two times. The screen will now read “05:00”.
7. Press the ENTER key one time to lock in the backup time.
8. Press the MODE key one time to switch to dose entry mode.
9. Press the ▲ key five times. The screen will now read “0005”.
10. Press the ◄ key until 5000 is displayed if checking a UVB device or 50.0 is displayed if
checking a UVA device.
11. Press the ENTER key one time to lock in the dose.
12. Press the ENTER key a second time to begin the treatment. The lamps will now come
on. When measuring ultraviolet light, be sure to apply SPF 30 sunscreen to exposed areas
of the skin and wear goggles while monitoring the device.

Table of Contents

Other manuals for Daavlin 1 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin 1 Series and is the answer not in the manual?

Daavlin 1 Series Specifications

General IconGeneral
BrandDaavlin
Model1 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals