MNL-00007 [3] Proprietary and Confidential Page 27 of 63
If power goes off during a treatment, the system will remember how much of the
treatment has elapsed. When power is restored, your device will be on “HOLD”. To
continue the treatment, press START. To cancel the treatment press END.
The small orange light above the display will be lit when a treatment is in process. It will
blink when the treatment is paused.
19.5 Flex Timer Range and Accuracy
Backup Time Range:
1 Second - 59 Minutes 59 Seconds (00:01 - 59:59)
Accuracy:
+ / - 5%
20.0 Flex Timer Instructions: 1 Series
20.1 Determining a Treatment Time
When using time as a dose, there are two treatment methods that can be used. The simplest is to
start at a prescribed time and then increase treatment times by a specific time increment or by a
percentage, per your physician’s protocol. Another method is to use the factory power output
levels (shown on the Power Output Certificate) to determine estimated “time equivalents”
needed to deliver doses in milliJoules (UVB) or Joules (UVA or visible light).
To determine a “time equivalent”, consult the laminated Time Chart that is provided and find the
power output on the top row (horizontal axis). The numbers in the column on the left (vertical
axis) represent a range of potential treatment doses. Find the dose then trace that row to the right
until it intersects with the column corresponding to the factory power output of the device. The
time equivalent is shown where the two lines intersect. For example, if the factory power output
is 4.0 mW/cm2 and the dose to be delivered is 280 mJ, then the appropriate treatment time will
be 01:10. See Table 5 Partial UVB Time Chart.
Note: For your convenience, and as an alternative to the UVB time chart, the latest version of
our free iPhone and Android application “Phototherapy Math” is available for download at the
iTunes app store and Android market. It quickly calculates treatment times, backup safety times,
and makes the calculation of dose and time increases simple. If using a PC, visit
www.phototherapymath.com for a web-based calculator and additional information.
mW/cm2
mJoules
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
7.0
240
01:36
01:20
01:09
01:00
00:53
00:48
00:44
00:40
00:34
280
01.52
01:33
01:20
01:10*
01:02
00:56
00:51
00:47
00:40
320
02.08
01:47
01:31
01:20
01:11
01:04
00:58
00:53
00:46
Table 5 Partial UVB Time Chart

Table of Contents

Other manuals for Daavlin 1 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin 1 Series and is the answer not in the manual?

Daavlin 1 Series Specifications

General IconGeneral
BrandDaavlin
Model1 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals