MNL-00007 [3] Proprietary and Confidential Page 43 of 63
light intensity will be much greater when treating directly on the acrylic surface (0 in) versus
treating from nine inches away. For this reason, it is necessary to select your desired treatment
distance before administering a treatment. By selecting your treatment distance, 0 in or 9 in, you
are informing the sensor how it should be evaluating and calculating the light output.
After unlocking your Integrated Dosimetry Control System (see section 14.0 Unlocking the
Device on page 15) you will need to select a treatment distance of either “0 in” or “9 in”.
Pressing the “POWER” button will allow you to toggle between these two options and pressing
“ENTER” will lock in your desired treatment distance.
25.5.2 Calibration Output Certificate
Here at Daavlin we measure the power output of each end every unit that we manufacture before
allowing it to leave our facilities. Included with your literature you will also find a “Calibration
Output Certificate” that contains your devices factory measured power output. Since your 1
Series has been designed to deliver treatments from two specific distances, 0 in and 9 in, your 1
Series will have two outputs noted in the “Current Value” box of your certificate. The higher of
the two output values corresponds to the “0 in” setting and the lower of the two values
corresponds to the “9 in” setting.
Note: When calculating your treatment backup time, be sure to use the correct output value for
the distance that you wish to treat from. More information about calculating backup times can be
found section 18.0 Flex Dosimeter Instructions: 1 Series on page 19.
25.6 Special Notes for 1 Series Units Equipped with the Digital Timer
25.6.1 Treatment Distances and Your Calibration Output Certificate
Your Daavlin 1 Series has been designed to administer treatments at two distinct treatment
distances. Treatments can be administered directly on the acrylic surface, making the treatment
of hands and feet comfortable and easy, or from a distance of nine inches away for conveniently
treating the face or other localized areas of the body.
Due to the inherent nature of phototherapy, the intensity of the ultraviolet light will be greater
when treating directly on the acrylic surface versus treating from nine inches away. In order to
insure accurate and effective treatments it is very important that you calculate your treatment
times using the power output that is appropriate to your desired treatment distance.
Here at Daavlin we measure the power output of each end every unit that we manufacture before
allowing it to leave our facilities. Included with your literature you will also find a “Calibration
Output Certificate” that contains your devices factory measured power output. Since your 1
Series has been designed to deliver treatments from two specific distances, directly on the acrylic
and nine inches away, you will have two outputs noted in the “Current Value” box of your
certificate. The higher of the two output measurements was taken from directly on the acrylic
surface while the lower of the two was measured from nine inches away.
Note: For hand and foot treatments, or any other treatment that will be administered directly on
the acrylic surface, always use the higher output value to calculate your treatment times. For