MNL-00007 [3] Proprietary and Confidential Page 26 of 63
19.2 Setting Up a Treatment Time
To administer a treatment, you must enter your desired treatment time. Based on the example
above the treatment time will be 1:10.
1. Press any key to awaken the unit. The word “CODE” will appear.
2. Enter in your key code. Unless you have changed it, the factory preset key code is 0007.
3. Once the key code has been entered, the display should read 00:00.
4. Press the ▲ key until the first digit of your desired time is displayed. If the treatment
time is one minute and ten seconds (01:10), the first digit will be a “1”.
5. Press the ◄ key to move the “1” one space to the left. The display should read 00:10.
6. Press the ▲ key until the next desired number is displayed (“1” in this case). The display
should read 00:11.
7. Press the ◄ key to move the “1” one space to the left. The display will now read 01:10.
8. When the desired time is displayed (in this case 01:10) press the ENTER key to lock in
your treatment time.
9. Put on your goggles and position yourself per the instructions.
10. Begin the treatment by pressing ENTER.
To repeat the same dose for another exposure, press the END key to clear the display. Then
press the MODE and POWER keys at the same time and release. This will automatically bring
up the last treatment administered. Press ENTER to start the treatment.
Note: The Digital Timer will beep three times to signal the end of a treatment. It will also flash
the treatment time on the display.
19.3 Special Functions of the Digital Timer
• To pause an active treatment press either the END or ENTER key.
• When a treatment is paused the screen will read “HOLD”. To resume a paused treatment,
press the ENTER key.
• To end or cancel an active treatment press the END key two times.
• If you make a mistake while entering a dose or backup time, repeatedly press the ◄ key
until all zeros are displayed.
• To check the age (in hours) of your lamps press and release the MODE and ◄ keys
together at the same time. The lamps must be off to do this.
19.4 Special Notes
The maximum time that can be entered is 59 minutes and 59 seconds
• It is not possible to enter 60 seconds. You must enter 1:00 minute instead.
• A short beep is sounded for every valid key entered. A long beep is sounded when an
invalid key is entered.