MNL-00007 [3] Proprietary and Confidential Page 29 of 63
It is not possible to enter 60 seconds. You must enter 1:00 minute instead.
A short beep is sounded for every valid key entered. A long beep is sounded when an
invalid key is entered.
If power goes off during a treatment, the system will remember how much of the
treatment has elapsed. When power is restored, your device will be on “HOLD”. To
continue the treatment, press START. To cancel the treatment press END.
The small orange light above the display will be lit when a treatment is in process. It will
blink when the treatment is paused.
20.5 Flex Timer Range and Accuracy
Backup Time Range:
1 Second - 59 Minutes 59 Seconds (00:01 - 59:59)
Accuracy:
+ / - 5%
21.0 UV Series X and UV Series T Information
21.1 Domestic and Non-CE Electrical Requirements
Daavlin offers the UV Series Phototherapy equipment in a variety of combinations. Units are
available with lamp quantities of 12, 16, 24, and 28 with various combinations of UVA,
Narrowband UVB, and Broadband UVB. While the type of ultraviolet lamp (UVA or UVB) does
not influence the electrical requirements, the number of lamps contained in a unit does determine
the voltage, amperage, wiring, and outlet necessary to safely support operation. The table and
diagrams below define these requirements for each model variation in the UV Series. Follow
these specifications and all national and local electrical codes when preparing the electrical
supply prior to the installation of a UV Series Device.
Note: This unit should be on a dedicated circuit.
Note: Daavlin UV Series X and UV Series T Phototherapy equipment contain no operator
replaceable parts.
NUMBER
OF LAMPS
VOLTS
HERTZ
AMPS
WIRE
GAUGE
OUTLET
DIAGRAM
OUTLET
12
110/120
60
15
14 AWG
A
NEMA No. 5-15R
16
110/120
60
15
12 AWG
A
NEMA No. 5-15R
24
110/120
60
30
10 AWG
C
NEMA No. L5-30R
28
110/120
60
30
10 AWG
C
NEMA No. L5-30R
12
220/240
50
15
14 AWG
Use Local Outlet
Use Local Outlet
16
220/240
50
15
14 AWG
Use Local Outlet
Use Local Outlet
24
220/240
50
20
12 AWG
Use Local Outlet
Use Local Outlet
28
220/240
50
30
10 AWG
Use Local Outlet
Use Local Outlet

Table of Contents

Other manuals for Daavlin 1 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin 1 Series and is the answer not in the manual?

Daavlin 1 Series Specifications

General IconGeneral
BrandDaavlin
Model1 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals