MNL-00007 [3] Proprietary and Confidential Page 24 of 63
1. Press the END key one time. This will clear the previous treatment and the display will
now read 00:00.
2. Press and release the MODE and POWER keys at the same time. This will reprogram
the backup time and dose used in your previous treatment.
3. Press the ENTER key to begin the treatment.
Note: If you have no additional treatments to administer or would like to program a treatment
utilizing a different dose or treatment distance, simply press the END key two times. This will
clear the display and return the unit to its energy saving mode.
18.5 Special Functions of the Integrated Dosimeter
To pause an active treatment press either the END or ENTER key.
When a treatment is paused the screen will read “HOLD”. To resume a paused treatment,
press the ENTER key.
When a treatment is in progress press the MODE key to toggle between watching the
dose administered count up or the treatment time count down.
To check the power output while a treatment is in progress press the POWER key.
To end or cancel an active treatment press the END key two times.
If you make a mistake while entering a dose or backup time, repeatedly press the ◄ key
until all zeros are displayed.
To check the age (in hours) of your lamps press and release the MODE and ◄ keys
together at the same time. The lamps must be off to do this.
18.6 Special Notes
It is not possible to enter a backup time of 60 seconds. You must enter 1 minute instead.
A short beep is sounded for every valid key entered. A long beep is sounded when an
invalid key is entered.
If power goes off during a treatment, the system will remember how much of the
treatment has elapsed. When power is restored, your device will be on “HOLD”. To
continue the treatment, press START. To cancel the treatment press END.
The small orange light above the display will be lit when a treatment is in process. It will
blink when the treatment is paused.
18.7 Flex Dosimeter Range and Accuracy
Backup Time Range:
1 Second - 59 Minutes 59 Seconds (00:01 - 59:59)
Dose Range:
UVA: 0.1 Joules - 999.8 Joules UVB: 1 Millijoule - 9998 Millijoules
Accuracy:
+ / - 5%
Calibration:
1 year or every 100 hours of use

Table of Contents

Other manuals for Daavlin 1 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin 1 Series and is the answer not in the manual?

Daavlin 1 Series Specifications

General IconGeneral
BrandDaavlin
Model1 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals