The UVA Time Chart (not shown here) is configured the same as the UVB Time Chart however
the doses are listed in Joules and the energy output levels and times are different than the UVB
output levels. Each chart is clearly marked.
20.2 Setting Up a Treatment Time
To administer a treatment, you must enter your desired treatment time. Based on the example
above the treatment time will be 1:10.
1. Press any key to awaken the unit. The word “CODE” will appear.
2. Enter in your key code. Unless you have changed it, the factory preset key code is 0007.
3. Once the key code has been entered, the display should read 00:00.
4. Press the ▲ key until the first digit of your desired time is displayed. If the treatment
time is one minute and ten seconds (01:10), the first digit will be a “1”.
5. Press the ◄ key to move the “1” one space to the left. The display should read 00:10.
6. Press the ▲ key until the next desired number is displayed (“1” in this case). The display
should read 00:11.
7. Press the ◄ key to move the “1” one space to the left. The display will now read 01:10.
8. When the desired time is displayed (in this case 01:10) press the ENTER key to lock in
your treatment time.
9. Put on your goggles and position yourself per the instructions.
10. Begin the treatment by pressing ENTER.
To repeat the same treatment time for another exposure, press the END key to clear the display.
Press any key to reawaken your unit and reenter your key code. Then press the MODE and
POWER keys at the same time and release. This will automatically bring up the last treatment
administered. Press ENTER to start the treatment.
Note: The Digital Timer will beep three times to signal the end of a treatment. It will also flash
the treatment time on the display.
20.3 Special Functions of the Digital Timer
To pause an active treatment press either the END or ENTER key.
When a treatment is paused the screen will read “HOLD”. To resume a paused treatment,
press the ENTER key.
To end or cancel an active treatment press the END key two times.
If you make a mistake while entering a dose or backup time, repeatedly press the ◄ key
until all zeros are displayed.
To check the age (in hours) of your lamps press and release the MODE and ◄ keys
together at the same time. The lamps must be off to do this.
20.4 Special Notes
The maximum time that can be entered is 59 minutes and 59 seconds

Table of Contents

Other manuals for Daavlin 1 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Daavlin 1 Series and is the answer not in the manual?

Daavlin 1 Series Specifications

General IconGeneral
BrandDaavlin
Model1 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals