EasyManua.ls Logo

DAKEWE SurePrint S200 - Page 79

DAKEWE SurePrint S200
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VII. Abnormalities and Troubleshooting
Dakewe (Shenzhen) Medical Equipment Co., Ltd. 73
www.dakewemedical.com
Figure 7- 8 Movement of the transfer push rod
Figure 7- 9Movement slides
(3) Gently pull forward the platform base, the slide falls off and take it out (Figure 7-10);
(4) Click “Reset” to reset the instrument.
3.Pull out the input hopper, and check whether too many slides are placed in the input hopper till
beyond the uppermost scale (Figure 7-11). If so, reduce the count of slides in the input hopper below the
scale mark.
Figure 7- 10 On-demand output unlocking
Figure 7- 11 Scale mark in the input hopper
4. Check if the front-back height difference in the cumulative thickness of slide in the input hopper is
greater than 1.5 mm; if so, reduce the count of the slides until normal print.
5. Check whether the slides in the input hopper are abnormal:
(1) Check whether the slides in the input hopper adhere to each other; if so, separate or take out the
adhered slides;
(2) Check whether the slide edge is broken; if so, take out the slide with broken edges (Figure 7-12);
Figure 7- 12Slide with broken edges
(3) Check whether the slides are regularly shaped; if there are irregularly shaped slides, take out the
them (Figure 7-13);