EasyManua.ls Logo

DAKEWE SurePrint S200 - Page 81

DAKEWE SurePrint S200
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VII. Abnormalities and Troubleshooting
Dakewe (Shenzhen) Medical Equipment Co., Ltd. 75
www.dakewemedical.com
Figure 7- 17 Print platform
Figure 7- 18 Input hopper holder
5. Stop immediately during print
Abnormality Description
The instrument immediately stops print, with a pop-up prompt.
Cause analysis
The maintenance door is opened during print.
Resolution
Close the maintenance door and the instrument will reset automatically.
6. The preset options in the drop-down box in the main interface are abnormal
Abnormality Description
The preset options in the drop-down box such as Prefix/Suffix/Custom1/Custom2 in the main interface
are set abnormally, and cannot be displayed in different lines.
Cause analysis
The separator entered is the full width comma (Figure 7-19).
Resolution
Enter the half width comma in the preset option box as the separator (Figure 7-20).
Figure 7- 19 Improperly using full width comma as
separator
Figure 7- 20 Preset of English comma