44 • Datacolor 650/600/400 Instrument Specifications Datacolor 650/600/400 User's Guide
Datacolor 650/600/400 Instrument Specifications
Specifications are subject to change without notice.
ITEM DESCRIPTION
Instrument Type
Dual beam integrating sphere with xenon flash lamp.
Measuring Geometry
Diffuse illumination, 8º viewing in conformance with CIE
publication No. 15.2 Colorimetry.
Illumination Source
Pulsed xenon, filtered to approximate D65.
Sphere Diameter
152 mm / 6.0 inches
Specular Port
Automated specular included or specular excluded
Spectral Analyzer
Proprietary SP 2000 analyzer with dual 256-diode array and
high-resolution holographic grating.
Wavelength Range
360 – 700nm
Photometric Range
0 to 200%
Black Trap
High performance
Aperture Configuration
Large Area View. 30mm illuminated/ 26mm viewed
Medium Area View. 20mm illuminated/ 16mm viewed
Small Area View. 9mm illuminated/ 5mm viewed
Ultra-Small Area View. 6.5mm illuminated/ 2.5mm viewed
X-Ultra Small Area View*. 3mm illuminated/ 2.5mm viewed
Power
85 to 264 VAC, 47 to 63 Hz, 80 VA peak, 35 VA typical
Absolute Operating
Environment
5º to 40º C, 5% to-85% RH, non-condensing
Interface
Serial: RS232, 9600/115200 baud (shipped as 19200)
USB: 1.1 or higher
Width
12-5/16 inches 312 mm
* Sample translucency may affect inter-lens agreement
** Dimensions shown are for horizontal mount instruments. The dimensions for the Datacolor 600V and Datacolor
44V (vertical mounts) can be found in Appendix, Dimension Comparison Vertical vs. Horizontal Mount.