Rooster Sensor100 USER MANUAL
Rooster
TM
Sensor100 USER MANUAL 62320MN000-A04 22
Upgrade From
Firmware Version
Upgrade Procedure
Calibration
Maintained?
User Settings Maintained?
V1.06.xx
Soft Reboot from Info
Screen
Yes
Yes
V1.04.xx
Power Cycle Procedure
No
Yes, if the Export EEPROM
function is used to save
Rooster settings before
upgrading firmware, and the
Import EEPROM function is
used to restore user settings
post upgrade.
V1.03.xx and
earlier
Power Cycle Procedure
No
No
You will need:
1. USB thumb drive with new firmware “Rooster.bin”
2. Rooster
TM
Sensor100
Review the firmware update procedure carefully before you begin. The firmware update
procedure must be allowed to complete without interruption. If the update procedure is
interrupted by user interference, a power failure, or some other circumstance, the existing
firmware may be corrupted, and the device may become nonfunctional. Contact Degree
Controls if your Rooster
TM
is inoperative as a result of an interruption during the firmware
update procedure.
Soft Reboot from Info Screen
Procedure:
1. Insert the USB thumb drive into the left side of Rooster
TM
.
2. Navigate to the Info [ ] screen and press the Upgrade [ ] button at the bottom of
the screen. Rooster™ will prompt you to confirm you wish to backup current settings and
upgrade the firmware.
3. Select “Yes”. Rooster™ will restart to read the USB thumb drive, and the Alarm LED turns
on. Once the upgrade process is complete, Rooster™ will automatically restart.
4. Navigate to the Info screen to validate that the new firmware version was properly
installed. The firmware release information listed next to “firmware” - should have
changed to the most current build. If this is not the case, repeat the upgrade procedure.
Power Cycle Procedure
Procedure: