EasyManua.ls Logo

Delta VFD-EL-W Series - Page 70

Delta VFD-EL-W Series
217 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 ParametersVFD-EL-W
65
To write all parameters, set Pr.00.02 = 0.
7: Long press the ENTER key for 5 sec. to lock the keypad and setting knob. When the
frequency command source is the keypad potentiometer (Pr.02.00 = 4), set Pr.00.02 = 7 after
setting the required frequency command, then the keypad potentiometer does not change the
drive’s frequency command.
8: Long press the ENTER key for 5 sec. to lock the keypad. Long press ENTER key for 5 sec.
again to unlock the keypad.
00.03
Start-up Display
Default: 0
Settings
0: F (frequency command)
1: H (output frequency)
2: A (output current)
3: U (user-defined)
4: FWD/REV command
Determines the start-up display page after power is applied to the drive.
Gets into the self-check state first when the drive starts-up, after displays “Ponand flashes for 5
sec., the drive turns to start-up page.
00.04
Content of Multi-function Display (User-Defined)
Default: 0
Settings
0: Display the content of user-defined unit
1: Display counter value (c)
2: Display the status of multi-function input terminal (d)
For example
Only MI1 terminal activates:
Only MI2 terminal activates:
MI2, MI3 and MI4 operate at the same time:
By analogy, MI1–MI4 are displayed in order from left to right
3: Display the drive's DC bus voltage (u)
4: Display the drive’s output voltage (E)
5: Display PID analog feedback signal (b)
6: Display the drive's power factor angle (n)
7: Display the drive’s output power (P)
8: Display the setting value and the feedback of PID control (P)
9: Display AVI analog input terminal signal (V) (I)
10: Display ACI analog input terminal signal (mA/V) (i)
11: Display IGBT temperature (°C) (h)
When Pr.00.03 is set to 3, use Pr.00.04 to select the displayed content as needed.
When Pr.00.04 = 5, the displayed PID feedback value is the percentage (%) of the terminal
measurement range.

Other manuals for Delta VFD-EL-W Series

Related product manuals